The reported issue was confirmed.The root cause of the reported issue is that the optical sensor is not situated correctly in the housing causing it to read incorrectly.The device was evaluated upon receipt.Optical sensor constantly reading "actuated".The manufacturing was updated as part.Applied gasket foam to optical sensor.Completed final functional.The device did not meet specifications, and was influenced by the reported failure.The device was not in use on a patient.The dhr is not required as the device has undergone previous servicing and therefore the reported event is not manufacturing related.Based on the results of this investigations, no additional action is required at this time.The instructions for use were found adequate and state the following: "1.2 indications for use indications: the bd sensica¿ urine output system is an automated system for continuous monitoring of urine output (uo) and core bladder temperature when connected to a temperature-sensing foley catheter.Target population: the bd sensica¿ urine output system is intended for bedside monitoring of urine output for any patient with an indwelling, urological catheter, drainage tubing and collection bag, typically in critical care settings or where close monitoring of urine output is desired.The bd sensica¿ urine output system is also intended for monitoring core bladder temperature when used with the bd sensica¿ temperature monitor module and a bd 400 series temperature-sensing foley catheter or equivalent.1.3 contraindications there are no known contraindications for use with patients who have indwelling, urological catheters in place.1.4 warnings general warnings medical electrical equipment requires special precautions regarding electromagnetic compatibility (emc) and needs to be installed and put into service according to the emc information provided in the charts at the end of these instructions for use (appendix a).Portable and mobile radiofrequency (rf) communications equipment can affect medical electrical equipment.The bd sensica¿ urine output system should not be stacked with other equipment.The bd sensica¿ urine output system has a degree of protection against electric shock of applied parts classified as type bf.This device is classified as an iec class i device.When using class 1 xp power model number: alm65us12 power supply, to avoid the risk of electric shock, this equipment must only be connected to a supply main with protective earth.Do not immerse or submerge the bd sensica¿ urine output stand, display monitor, ring, temperature monitor module, temperature sensor cables, or monoplug adapter in water or other liquids.Do not pour liquids over the display monitor, ring, temperature monitor, cables or monoplug adapter.If liquids accidentally spill onto the device(s), wipe off liquid with soft cloth as soon as possible.See section 14 for complete care instructions.This device is not suitable for use in the presence of flammable mixtures.This device is not suitable for use in oxygen rich environments.Operation warnings federal law restricts this device to sale or use by or on the order of a licensed physician.The bd sensica¿ urine output system should only be used on the request by and under the direction of a licensed physician.Do not use bd sensica¿ ring (disposable) if product appears broken.The ¿accuracy check¿ should be performed prior to first use, when there is a question about system accuracy, and/or after any servicing or repairs to ensure accurate measurements.Always use a new bd sensica¿ ring to complete an accuracy check.If the accuracy check fails call bd customer service at (b)(6).This self-test takes about one minute and can be accessed from the new patient and settings screen or from the administrator settings.Do not change time and date settings during system operation.Time and date settings should be set when system is initially installed or between monitoring sessions, if needed.Ensure that the urine drainage bag hanging on the bd sensica¿ ring is always below the level of the patient¿s bladder to maintain continuous, gravity flow of urine into the collection bag.When connecting the single patient use bd sensica¿ ring to the system stand, use a firm, clockwise twisting motion.In order to avoid damaging components, do not apply excessive force or torque to the ring or the system¿s ring interface when connecting the device.To properly stabilize the system for accurate urine output measurement precautions: during system start up and in general practice, plug the bd sensica¿ urine output system into a wall power supply whenever possible.The system screen will dim when the system is unplugged to maximize battery life.After using the system on battery back-up, plug it back into the wall power supply for recharging and to avoid system shut down due to a drained battery.After removing a patient from urine output monitoring with the bd sensica¿ urine output system, shut down the system completely to avoid battery drainage while not in use.Low priority alarms may interrupt urine output monitoring and require user action to resume normal urine output monitoring functions.If the power cord icon is displayed on the screen, the power supply is not connected.If the system is repositioned, always ensure that the power supply is fully seated on both ends.Do not disturb the system during stabilization." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|