Catalog Number 2C8750 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was a separation issue during the use of clearlink system y-type blood/solution set.The separation issue was further described as, ¿they found that the end of the intravenous (iv) tubing had snapped off into the patient's iv clave¿.This event occurred upon disconnecting the iv fluids from the patient.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection with the naked eye was performed and did not identify any abnormalities that could have contributed to the reported condition.Pressure testing was performed, and no leak was identified in the set.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|