ABBOTT MEDICAL G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Model Number SGC0701
Device Problems Positioning Failure (1158); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2023
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned for evaluation as the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report inability to straighten it was reported this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.It was noted the patient a severely tortuous iliac vein leading into the inferior vena ceva (ivc).The steerable guide catheter (sgc) was inserted, but when applying +/- knob, the sgc was unable to curve or straighten.Therefore, the sgc was removed.Upon removal, it was observed the shaft was deformed and the tip was kinked.No additional devices were attempted and the procedure was discontinued.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation determined the reported unable to curve, unable to straighten, difficult to advance, and deformation due to compressive stress of the sgc appear to be related to patient morphology/pathology.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16605267
MDR Text Key311975067
Report Number2135147-2023-01324
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/23/2023
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number21019R2037
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/02/2023
Initial Date FDA Received03/23/2023
Supplement Dates Manufacturer Received03/28/2023
Supplement Dates FDA Received04/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age89 YR
Patient SexFemale
-
-