Initial reporter address 1: (b)(6).Device evaluated by mfr.: returned product consisted of the solent omni thrombectomy.The pump, effluent/supply line, shaft, tip, piston, and spike line were visually inspected.Visual examination revealed no damages.Microscopic examination revealed that the hypotube is separated 21.5cm from the tip.Inspection of the device presented no other damage or irregularities.
|
Reportable based on device analysis completed on 02-march-2023.It was reported that an error message was displayed.The 70% target lesion was located in the non-tortuous and severely calcified of left lower limb deep vein.An angiojet solent omni catheter was selected for thrombectomy procedure.However, during the procedure under thrombectomy mode, the system alerted a check saline error message and there was fluid observed in the pump.The procedure was completed with another of same device.There were no patient complications were reported and the patient's status was stable.However, returned device analysis revealed a separated hypotube.
|