C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE Back to Search Results
Model Number 777626
Device Problems Flaked (1246); Device Contamination with Chemical or Other Material (2944)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 03/02/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that white grease was found upon unpacking the ureteral stent.The surgery time was prolonged and surgery schedule by health care professional was affected.
 
Manufacturer Narrative
The reported event was confirmed and the cause was unknown.Although a specific cause could not be determined, potential root cause for this event could be inadequate material selection, and/or bonding between layers, degradation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "inspect all guidewires for damage prior to use.Bending or kinking during or prior to placement could damage the guidewire.Do not attempt to use the guidewire if it has been damaged.Use of a damaged wire may result in damage to the urinary tract." "do not reshape the guidewire by any means.Attempting to reshape the guidewire may cause damage resulting in release of fragments into the urinary tract.Failure to exercise proper caution may result in damage to the urinary tract." "do not manipulate or remove the guidewire through a metal cannula or needle.This may result in destruction/separation of the outer jacket of the wire requiring retrieval." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that white grease was found upon unpacking the ureteral stent.The surgery time was prolonged and surgery schedule by health care professional was affected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Type of Device
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16617987
MDR Text Key312046601
Report Number1018233-2023-02034
Device Sequence Number1
Product Code FAD
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K983498
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number777626
Device Catalogue Number777626
Device Lot NumberNGGV2068
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/02/2023
Initial Date FDA Received03/27/2023
Supplement Dates Manufacturer Received06/26/2023
Supplement Dates FDA Received07/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-