SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA1700/000/000JP |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the during the pre-use check, leakage of air from the product was observed.No patient injury or clinical affects was reported.
|
|
Manufacturer Narrative
|
Month and year of event have been provided, day is unknown.Lot number, expiration date and udi number.Device manufacture date is unknown.No product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
One device was received.As a result of visually observing the device, a crack on the connector was found.A leak test was conducted, and the complaint was confirmed as the circuit tube was leaking due elbow damage.Based on the analysis conducted in the sample provided, the root cause could be due to the improper use of the product.A device history record (dhr) review could not be performed as the lot number was unknown.No action taken, this failure condition will be monitored for threshold or escalation.
|
|