The initial reporter stated they received discrepant results for one patient sample tested with the elecsys total psa assay on a cobas e 801 module, serial number (b)(4).The sample initially resulted in a total psa value of 0.19 ug/l.The result was deemed incorrect.On (b)(6) 2023, the sample was repeated two times, resulting in values of 15.8 ug/l and 15 ug/l.On (b)(6) 2023, the sample was repeated two times, resulting in values of 14.8 ug/l and 15.4 ug/l.No questionable result was reported outside of the laboratory.
|