Model Number OER-ELITE
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2023
Event Type  malfunction  
Manufacturer Narrative
An olympus field service engineer (fse) was dispatched to evaluate and repair the device.The fse removed and replaced the missing connector label and the purple connector, the unit was repaired and verified according to the original equipment manufacturer (oem) instructions, software attributes have been verified and confirmed.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus, the endoscope reprocessor was missing a connector label and its purple connector was damaged.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, it is likely the event occurred due to when the user used the equipment, they forgot to attach the valve during connection of the connecting tube again after disconnected it from reprocessing basin.However, a definitive root cause could not be determined.The following information is stated in the instructions for use (ifu): ¿5.6 inspecting the connectors check the following for each connector.·the connector should be fixed firmly.The o-rings should be free of irregularities such as cracks, tears, or dents.If any irregularity is found, do not use the reprocessor and contact olympus.Warning do not use the reprocessor if any connector seems to be damaged or defective.Using the reprocessor when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral devices or facilities near the equipment.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16628821
MDR Text Key312419585
Report Number9610595-2023-05150
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170404047
UDI-Public04953170404047
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201920
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-ELITE
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/28/2023
Initial Date FDA Received03/28/2023
Supplement Dates Manufacturer Received06/26/2023
Supplement Dates FDA Received07/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-