Model Number 0998-00-3013-53
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2023
Event Type  malfunction  
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) has an electrical failure alarm.There was no patient involvement reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
N/a.
 
Event Description
N/a.
 
Manufacturer Narrative
Additional contact information: (b)(6).A getinge field service engineer (fse) evaluated the unit for the reported malfunction.The fse replaced the front end pcb.The fse performed a unit checkout.The unit passed all functional and safety tests.The iabp was cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16635065
MDR Text Key312426771
Report Number2249723-2023-01669
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107394
UDI-Public10607567107394
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3013-53
Device Catalogue Number0998-00-3013-53
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/17/2023
Initial Date FDA Received03/29/2023
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/31/2024
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-