The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on july 14, 2020.A sample was not received for the investigation.Because a sample was not returned, we were unable to perform functional and visual evaluations to confirm the reported condition and determine the root cause.However, a corrective and preventative action has been initiated to address the reported condition.
|
The customer reported that on (b)(6), in the 2nd department of general surgery of (b)(6) cancer hospital, an abnormality occurred after the patient was intubated, and the patient was immediately sent for x-ray confirmation.It was found that the gravity head of the nasal feeding tube was broken and fell into the patient's airway.The hospital immediately treated the patient.At present, the patient is in normal condition, but emotional.Per additional information provided on (b)(6), 2023 the weighted tip detached during the placement of the ng tube.The patient did not display any symptoms.The mention of the patient being intubated was in reference to the placement of the ng tube.The weighted tip was removed by surgical procedure.A new ng tube was then placed.
|