Model Number 45038 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/08/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that foot pedal stuck occurred.An angiojet console was used for a thrombectomy procedure.However, when the foot pedal is pressed, the device continues to pulsate fluid into the patient.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: angiojet footswitch was received.A visual inspection was done on the footswitch, and it passed visual inspection.The received footswitch was installed onto the golden angiojet console.Then a functional test was performed.During the test, the console continued to pump even though there was no pressure applied to the footswitch.The complaint was confirmed.
|
|
Event Description
|
It was reported that foot pedal stuck occurred.An angiojet console was used for a thrombectomy procedure.However, when the foot pedal is pressed, the device continues to pulsate fluid into the patient.There were no patient complications reported.
|
|