• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIKO AB GOLVO 9000; LIFT, PATIENT, NON-AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIKO AB GOLVO 9000; LIFT, PATIENT, NON-AC-POWERED Back to Search Results
Model Number 2000045
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2023
Event Type  malfunction  
Event Description
The customer alleged that sparks occurred when the power cable was connected to a socket.There was no allegation of patient or caregiver injury or death reported from this alleged incident.This report was filed in our complaint handling system as (b)(4).
 
Manufacturer Narrative
The electrical parts of hillrom lifts are compliant with iec 60601-1 (medical electrical equipment - part 1: general requirements for basic safety and essential performance) which applies to the basic safety and essential performance of medical electric equipment and medical electric systems.Even though all hillrom lifts comply to the above mentioned electrical standard, there is still a risk that abnormal wear and tear can damage the cables.Therefore, hillrom states in the periodic inspection manual for liko mobile lifts (3en371001 rev.5) under section 8: "verify that all cables are properly inserted.Re-insert if uncertain.Verify battery charge by inspecting the charger indicator.Check cables and connectors for damage or wear." in the instruction manual for golvo 9000 lifts (7en140108 rev.4) states, under inspection and maintenance section: "a periodic inspection of the lift should be carried out at least once per year.Periodic inspection, repair and maintenance may be performed only in accordance with the liko service manual by personnel authorized by hillrom and using original liko spare parts." ¿for trouble-free use, certain details should be checked each day the lift is used: inspect the lift and check to make sure that there is no external damage.¿ if the instruction as outlined above are followed it is unlikely for an injury to occur due to this error.The charging cable was replaced to resolve the issue, the lift is functioning as designed.In conclusion, although the reported event did not result in a patient injury, the report of a charging cable sparking during charge could contribute to a serious injury or death, if the malfunction were to recur.Therefore, hillrom considers this complaint a reportable malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOLVO 9000
Type of Device
LIFT, PATIENT, NON-AC-POWERED
Manufacturer (Section D)
LIKO AB
nedre vagen 100
lulea, norrbottens lan [se-25] 975 9 2
SW  975 92
Manufacturer Contact
antonio acquafredda
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key16654870
MDR Text Key313194314
Report Number8030916-2023-00026
Device Sequence Number1
Product Code FSA
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number2000045
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/16/2023
Initial Date FDA Received03/31/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-