|
Model Number 24674 |
Device Problems
Break (1069); Material Rupture (1546)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/24/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that balloon rupture and shaft break occurred.The target lesion was located in the moderately tortuous and moderately calcified arteriovenous fistula.A 12.0 x 80, 75cm mustang balloon catheter was advanced for dilation.However, after the balloon was inflated more than three times at above 20 atmospheres, the balloon burst.It was also noted that the middle of the catheter was separated outside the patient.The device was simply pulled out, and the procedure was completed with another of the same device.No patient complications were reported, and the patient condition was stable.
|
|
Event Description
|
It was reported that balloon rupture and shaft break occurred.The target lesion was located in the moderately tortuous and moderately calcified arteriovenous fistula.A 12.0 x 80, 75cm mustang balloon catheter was advanced for dilation.However, after the balloon was inflated more than three times at above 20 atmospheres, the balloon burst.It was also noted that the middle of the catheter was separated outside the patient.The device was simply pulled out, and the procedure was completed with another of the same device.No patient complications were reported, and the patient condition was stable.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: mustang 12.0 x 80, 75cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A complete balloon circumferential tear was identified located approximately 50mm distal of the proximal balloon bond.This type of damage most probably occurred when the device was being withdrawn through the sheath.The distal section of balloon material including a section of shaft, the distal markerband and tip of the device was not returned for analysis.The rated burst pressure for this device as per mustang specification is 14 atmospheres.A visual and tactile examination found the shaft of the device to have completely separated at a location proximal of the distal tip bond.The distal section of shaft including the distal section of balloon material, the distal markerband and tip of the device was not returned for analysis.The shaft is also severely stretched distal of the proximal markerband.The distal section of the device including a section of balloon material, a section of shaft the distal markerband and tip of the device was not returned for analysis.A visual examination observed that the distal section of the device including the distal markerband, a section of balloon material, a section of shaft and tip of the device was not returned for analysis.The proximal markerband was undamaged in the correct position on the device.
|
|
Search Alerts/Recalls
|
|
|