• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-HQ190
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, air/water cylinder discolored, suction cylinder discolored, switch box discolored, right/left discolored, up/down knob discolored, grip scratched, scope cover case unit scratched, due to clogging of nozzle, water removal ability did not meet the standard value, due to wear of angle wire, the play of up/down/right/left knob out of the standard value, distal end cover chipped, objective lens cracked, light guide lens scratched, light guide lens cracked, connecting tube coating peeling, and universal cord coating peeling.The faulty parts were replaced, and the device was returned to the user facility.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
A user facility returned the olympus asset, evis exera iii gastrointestinal videoscope to the olympus service center.Upon inspection and testing of the returned device, the scope air/water tube was found clogged with foreign material due to incorrect reprocessing.This report is being submitted for the event found during evaluation.The customer did not report any problems associated with the device and no patient involvement, no harm or user injury reported to olympus.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The foreign material was unable to be identified.There have been no reports of deformation or damage to the air/water channels.The detection method is described in the instruction manual gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection and prevention measures are described in the instruction manual gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16656122
MDR Text Key312638706
Report Number9610595-2023-05383
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170305276
UDI-Public04953170305276
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-HQ190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/05/2023
Initial Date FDA Received03/31/2023
Supplement Dates Manufacturer Received06/07/2023
Supplement Dates FDA Received07/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-