Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation: it was reported by a representative of (b)(6) medical center ((b)(6), usa) that on (b)(6) 2023 an ultrathane mac-loc locking loop multipurpose drainage catheter (rpn: ult10.2-38-25-p-6s-clm-rh; lot#: 15165300) separated.The device was placed during an exchange procedure and was inserted over a wire guide into the gallbladder.The catheter was attached to an extension tube and drainage bag, and the fittings were noted to be securely attached.About 10 minutes after placement, the catheter separated near the mac-loc hub.As a result, the patient required an additional procedure to remove and replace the device.Reviews of documentation including the complaint history, device history record (dhr), quality control procedures, manufacturing instructions (mi), and instructions for use (ifu), as well as a visual inspection of the returned device, were conducted during the investigation.One catheter was returned in a used and damaged condition to cook for evaluation.The mac-loc adaptor was received separated from the shaft of the catheter.The point of separation appears to have been just below the mac-loc adaptor.Additionally, the proximal end of the shaft at the location of separation exhibited jaggedness and was missing the flare.Upon disassembling of the cap from the mac-loc adaptor, absence of the flare was noted.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot 15165300 and the related subassembly lots revealed no related non-conformances.A review of complaint history for this lot found two additional complaints from the same customer for the same failure.To date, there have been no additional complaints received for leakage or hub separation from these lots.Additional related lots manufactured in a similar time period and lots made by the relevant operator were reviewed.No additional related complaints were noted from these lots.Some lots showed relevant nonconformances, but all nonconforming devices were scrapped prior to release.Since there are no related additional complaints received from the related lots, there is no evidence that additional nonconforming product exists in house or in the field cook also reviewed product labeling.The product ifu, [t_multi2_rev1] ¿multipurpose drainage catheter,¿ provides the following information to the user related to the reported failure mode: how supplied upon removal from package, inspect the product to ensure no damage has occurred.The information provided upon review of the dmr, ifu, dhr, and analysis of the returned device suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, inspection of the returned device, and the results of the investigation, cook medical has concluded the cause can be traced to component failure, without any design or manufacturing issue.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|