• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART MRX -EMS DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART MRX -EMS DEFIBRILLATOR Back to Search Results
Model Number M3536A
Device Problem Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2023
Event Type  malfunction  
Event Description
This report is based on information provided by philips remote service engineer (rse) and has been investigated by the philips complaint handling team.Philips received a complaint on philips heartstart mrx defibrillator indicating that the power module was allegedly malfunctioning, and not functioning properly.There was reportedly no patient involvement.Upon remote investigation, it was found that the ac power module was faulty and needed to be replaced.However, the ac power module field replacement kit was on backorder, causing a delay in the repair of the device.This complaint investigation has been impacted by logistical limitations associated with the service parts supply chain (sps) returns process.This limitation is beyond the immediate control of the complaint handling group.No further investigation of the component is possible until the limitations are resolved.Based on the information available and the testing conducted, the cause of the reported problem was a malfunction of the ac power module.Based on the information available and results of additional analysis, no further action is necessary at this time.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.Philips recommended replacing the ac power module field replacement kit to resolve the issue.It has been concluded that no further action is required at this time.If additional information is received, the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX -EMS DEFIBRILLATOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
allison prater
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key16657429
MDR Text Key312460420
Report Number3030677-2023-01356
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838006652
UDI-Public00884838006652
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Device Catalogue NumberM3536A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/02/2023
Initial Date FDA Received03/31/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-