Model Number 90620US
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2023
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the cord leading to the battery pack of the duo headlight 2 bay system - us (90620us) has a short in it.No patient/user injury, death or surgical delay occurred as a result of the event.
 
Manufacturer Narrative
The reported duo headlight 2 bay system - us (90620us) was not returned for investigation; therefore, an evaluation of the device could not be performed and a definite root cause of the reported issue could not be identified.Based on the reported complaint, the probable root cause for this 90620us headlight having a short in the cord from the headband to the battery pack is damage to the cord caused by rough handling/environmental damage.The issue of a "short" refers to poor connection of the power pins that can potentially result in a failing connection to the battery port.No risk of harm would be incurred by this type of short, and the low voltage would not cause a hazard.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO HEADLIGHT 2 BAY SYSTEM - US
Type of Device
PFM05
Manufacturer (Section D)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16664715
MDR Text Key312865916
Report Number3006697299-2023-00034
Device Sequence Number1
Product Code FSR
UDI-Device Identifier10381780490746
UDI-Public10381780490746
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number90620US
Device Catalogue Number90620US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/09/2023
Initial Date FDA Received04/03/2023
Supplement Dates Manufacturer Received05/09/2023
Supplement Dates FDA Received05/19/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-