• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 87035
Device Problem Material Perforation (2205)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2022
Event Type  malfunction  
Manufacturer Narrative
The intellamap orion catheter was returned to boston scientific for analysis.Visual inspection revealed the missing locking knob and the distal section of the catheter including the cage/array was detached and not returned.The reported event was confirmed.
 
Event Description
Reportable based on analysis completed on 28mar2023.It was reported that during mapping and ablation procedure with the intellamap orion catheter, the steering/clutch knob was broken and detached from the catheter handle.The catheter was replaced, and the procedure was completed.There were no patient complications.Analysis of the returned device revealed that the distal portion of the device containing the basket array was severed and not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key16664739
MDR Text Key312594822
Report Number2124215-2023-15443
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K143481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/03/2024
Device Model Number87035
Device Catalogue Number87035
Device Lot Number0030077367
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/24/2023
Initial Date FDA Received04/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/02/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-