Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 03/14/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that during an open reduction and internal fixation of a left distal radius fracture, the screwdriver tip broke off inside the proximal locking screw.No intervention has been reported.The fractured piece has been retained in the patient body.No adverse event has been reported as a result of the malfunction.
|
|
Event Description
|
No further event information is available at the time of this report.
|
|
Manufacturer Narrative
|
Reported event was confirmed by review of pictures.Visual examination of the provided pictures identified that the tip of the screwdriver broke off.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|