Model Number GIF-HQ290 |
Device Problem
Contamination /Decontamination Problem (2895)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/05/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus that during maintenance of the evis lucera elite gastrointestinal videoscope, the scope failed leak testing.Upon inspection and testing of the returned device, foreign material was found protruding from the scope air/water channel due to incorrect reprocessing.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.There was no patient involvement reported with this event.
|
|
Manufacturer Narrative
|
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, scope connector corroded, fluid ingress in scope connector, image error code, biopsy channel leaking, up/down/right angle not to specification, and up/down/right/left angle play.The faulty parts were replaced, and the device was returned to the user facility.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, it is likely a leak occurred in the forceps channel, so reprocessing could not be completed and foreign matter remained in the air and water nozzles.According to the instructions for use (ifu), there is the following description to contact olympus when a leak is discovered.So it's not a departure from ifu to abort reprocessing when a leak occurs."reprocessing manual_ cleaning, disinfection and sterilization procedures_ leakage testing_ perform the leakage test caution:if you locate a leak,remove the endoscope from the water and contact olympus." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|