E1: establishment name: (b)(6) university.The subject device has been returned to olympus for evaluation.In addition to the reportable malfunction mentioned in b5, it was observed, the bending cover was found to be leaking, and the lock engagement (fe lever) was stuck.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
The customer reported to olympus, during reprocessing, the uretero-reno videoscope was leaking.Upon inspection of the customer¿s returned device it was observed, the bending section could not be controlled due to corrosion on the angle mechanism and the metal skeleton inside the bending tube was protruding.This report is being submitted for the malfunction found during evaluation.There was no harm or user injury reported due to the event.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation for phenomenon one, it is likely that the angulation failed due to corrosion inside the bending section and control section caused by a leak at the bending section.Based on the results of the investigation for phenomenon two, it is likely that the cable support was broken and protruded from the bending section cover due to excessive force on the passive bending section.The event can be prevented by following the instructions for use which state: "·inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.·do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged." olympus will continue to monitor field performance for this device.
|