• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND SRD S5 ERC TUBING CLAMP/500MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND SRD S5 ERC TUBING CLAMP/500MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-05-60
Device Problems Difficult to Open or Close (2921); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2023
Event Type  malfunction  
Event Description
Livanova deutschland has received a report that the electrical remote-controlled tubing clamp (erc) is not working correctly during a procedure.There is no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the s5 erc tubing clamp.The incident occurred in united states.A livanova field service engineer was dispatched to the customer site and the erc was replaced.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
Through follow-up communication livanova learned that the involved electrical remote-controlled tubing clamp (erc) was not sensing tubing and making squealing sounds when clamping and unclamping.As per device instructions for use, the issue that the tubing is not recognized it means that the clamping lever is in the ¿close¿ position.Simultaneously, an alert tone sounds at the scp system.Even in the case the clamp remains in closed position, it can always be manually opened.Automatic clamps devices are equipped with a handle aimed at the manual opening of the clamp releasing the engaged tube.Therefore, in case of undesired closure of the automatic clamps occurs, opening it is always possible to act on the handle to prompt perform a manual opening which is unlikely to cause or contribute to death or serious injury.Therefore, the event has been re-assessed as not reportable.Based on the review of similar complaints and according to device instructions for use, the "no tubing recognized" error message can be displayed when: - no tube is inserted into the clamp; - the wrong tube is inserted into the clamp.Regarding the noise issue and considering that the claimed clamp was manufactured in 2022, the wearing of the unit ball bearing can be ruled out and most likely the noise has been caused by residues of dried fluids on the shafts, which which obstruct the movements generated noise.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SRD S5 ERC TUBING CLAMP/500MM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
arvada
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16681779
MDR Text Key312722092
Report Number9611109-2023-00148
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier04033817901051
UDI-Public(01)04033817901051(11)220531
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-05-60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/09/2023
Initial Date FDA Received04/05/2023
Supplement Dates Manufacturer Received05/19/2023
Supplement Dates FDA Received06/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-