Additional information added to d9, h3, h6 and h10 h10: the device was received for evaluation.During visual inspection, the dialysate pump segment was observed not glued in the support plate.The visual inspection of the set showed the dialysate pump segment is not glued to the support plate.The reported condition was verified.The cause of the condition was due to inadequate amount of solvent on the tubing during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|