Ge healthcare investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.No patient information provided to date.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.Device evaluation anticipated, but not yet begun.
|
The complaint for the unit not alarming when the patient's spo2 dropped below 50% is applicable only to the patient monitor that was in use.The avance does not measure, and therefore does not alarm for, blood oxygen levels (spo2).The avance has the capability to measure fraction inspired and fraction expired values for o2, co2, and agent.These values would only lead to an alarm based on the alarm limits set by the clinician.There was no allegation related to avance alarms or performance from the event.The ge healthcare fe inspected the anesthesia machine and found no issue.The patient monitor was inspected.The service personnel found that the alarms had been turned off.The staff was informed of what they found.The monitor was tested and no other issues were identified with respect to the alarm system.On follow-up, the customer acknowledged that the issue was due to user error and did not want any further investigation into it by ge healthcare.
|