Model Number T505U223 |
Device Problems
Gradient Increase (1270); Perivalvular Leak (1457); Material Separation (1562); Insufficient Information (3190)
|
Patient Problems
Aortic Valve Stenosis (1717); Pseudoaneurysm (2605); Valvular Insufficiency/ Regurgitation (4449); Insufficient Information (4580)
|
Event Date 03/29/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Medtronic received information that 10 years and 9 months post implant of this 23mm aortic bioprosthetic valve, it was explanted and replaced with a 27mm bioprosthetic valve.The reason for the replacement was not reported.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received that reported the reason for replacement as moderate to severe aortic paravalvular regurgitation, severe aortic stenosis and a aortic root pseudoaneurysm.It was stated that the valve was observed to be partially dehisced indicating the opening of a pseudoaneurysm.It was noted that there was a big pseudoaneurysm at the posterior aspect of the left coronary sinus and another pseudoaneurysm in the noncoronary sinus level.The valve and pseudoaneurysm were observed to be calcified.It was reported that there were two paravalvular openings, one going to the left coronary sinus and another to the noncoronary sinus.It was also stated that peak and mean transvalvular gradients of 85mmhg and 44mmhg.It was also reported that the visual left ventricular ejection fraction (lvef) was 40-45%, the max left ventricular outflow tract (lvot) was 1.16 m/s, the right ventricular systolic pressure was mildly increased, the left atrium was moderately dilated and the aortic valve area (ava) was 1.1 cm2.The patient experienced symptoms of shortness of breath.No additional adverse patient effects were reported.
|
|