• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Decrease in Suction (1146)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2023
Event Type  malfunction  
Manufacturer Narrative
Ge healthcare's investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.No report of patient involvement.E1: the initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.D4: unique identifier: (b)(4).Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.Device evaluation anticipated, but not yet begun.
 
Event Description
It was reported that there was a malfunction resulting in loss of suction.There was no patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The overflow safety trap was recommended for replacement to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH   214028
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key16699975
MDR Text Key312888763
Report Number9710602-2023-00337
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/09/2023
Initial Date FDA Received04/07/2023
Supplement Dates Manufacturer Received09/11/2023
Supplement Dates FDA Received09/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-