The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.The front socket was found loose due to the socket being loose internally and the slider switch not functioning properly.Additionally, as noted, a non-olympus lamp had been installed in the subject device.Heavy dust was noted inside the unit, and the fan air vent was found clogged.There was also minor rust noted and corrosion inside the front panel.If additional information becomes available following the device evaluation, a supplemental report will be filed.
|
E2, e3, & g2 - information has been obtained and provided.This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the loose socket could not be identified.Olympus will continue to monitor field performance for this device.
|