• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number B35200
Device Problem Electromagnetic Interference (1194)
Patient Problem Shaking/Tremors (2515)
Event Date 04/03/2023
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient's left side got more shaky about the last week or so.Prior to this event, the patient said they got an mri about 10 days or 2 weeks ago and stated the implantable neurostimulator (ins) was turned on during the mri and was not in mri mode because the mri technicians didn't say anything about it.Before calling, the patient increased stimulation for their left side and mentioned that their right side was fine, so they did not adjust the right side.The patient said they will monitor their left side for a couple of days to see if it gets better.
 
Event Description
Additional information received from the consumer reported the cause of the shaking was them turning stimulation off.The patient was admitted to the hospital for tia symptoms after they had the mri (which resolved itself in 24 hours).To resolve the shaking and potential mri interference the patient couldn¿t have any other mri¿s or turn stimulation off.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPT
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16715996
MDR Text Key313059462
Report Number3004209178-2023-04829
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00763000519216
UDI-Public00763000519216
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberB35200
Device Catalogue NumberB35200
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/10/2023
Initial Date FDA Received04/11/2023
Supplement Dates Manufacturer Received06/01/2023
Supplement Dates FDA Received06/06/2023
Date Device Manufactured09/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
Patient Weight67 KG
-
-