Model Number 3810002 |
Device Problems
Mechanical Problem (1384); Device Slipped (1584); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/17/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported that the iv pole does not stay elevated.The iv pole clamp is installed.No injury was reported for this incident, and no patient was connected at the time the iv pole was sliding down.Therefore, no patient information is reasonably known.
|
|
Manufacturer Narrative
|
Investigation is in process.A follow-up report will be provided.
|
|
Event Description
|
The customer reported that the iv pole does not stay elevated.The iv pole clamp is installed.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
|
|
Manufacturer Narrative
|
This report is being filed to provide additional information in h.6 and h.10.Investigation: a terumo bct service technician checked out the machine at the customer site and was able to confirm the reported condition.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Correction: a field service engineer adjusted the iv pole button.Root cause: a root cause assessment was performed for this complaint.The root cause was determined to be related the need for an iv pole button adjustment.
|
|
Search Alerts/Recalls
|