H10: h3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The insertion device was returned with a deformed and fractured anchor but no suture material.The proximal end of the insertion device has dents from impact.The inner shaft is bent approximately 90 degrees just above where it exits the outer shaft.The proximal end of the anchor is below the bend of the inner shaft and deformed.The distal end of the anchor is deformed.All returned items have debris on them.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A clinical review states one undated photo of the device was provided for review and confirms the reported breakage.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|