Catalog Number 100/860/075CZ
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2022
Event Type  malfunction  
Event Description
It was reported that a device arrived damaged.Discovered upon opening.No injury.
 
Manufacturer Narrative
H6: event problem and evaluation codes: updated.H10: a device history record dhr review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found a hole in cuff.Functional testing found that it is not even possible to inflate cuff as it leaks so badly.The root cause of the reported issue was found to be manufacturing.History of internal ncrs showed that there was raised internal nonconformity report due to cuff tear which was found after inflation test.Corrective actions have been taken including initiation of quality alert, training of whole production during production town hall meeting, replacement of gauges and equipment with sharp edges by blunt versions, warning in manufacturing procedure and establishing better organization on affected workplace.No trend of confirmed complaints in relation with this issue was identified, cuff tear prior use is reported rarely, by implementation of actions from the internal ncr we reduced likelihood of such failure.Reported lot number was manufactured may 2021 prior implementation of corrective actions december 2021.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE ULTRA TRACH TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ  753 01
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16725553
MDR Text Key313156343
Report Number3012307300-2023-04024
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K030570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number100/860/075CZ
Device Lot Number4134718
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/15/2022
Initial Date FDA Received04/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-