• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ESSENTIO MRI DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ESSENTIO MRI DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L111
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2023
Event Type  Injury  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that this pacemaker had entered safety mode.Device replacement and product return were recommended.This pacemaker remains in service at this time.No adverse patient effects were reported.
 
Manufacturer Narrative
The product has been received for analysis.This report will be updated upon completion of analysis.
 
Event Description
It was reported that this pacemaker had entered safety mode.Device replacement and product return were recommended.This pacemaker remains in service at this time.No adverse patient effects were reported.Additional information received reported that this pacemaker was explanted and replaced due to entering safety mode.No additional adverse patient effects were reported.This pacemaker was returned for analysis.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode.It was confirmed that critical therapy did not remain available.The device case was opened and the battery was removed and forwarded for detailed testing, which revealed no battery-related anomalies.Despite analysis, the root cause of the clinical observations could not be confirmed.
 
Event Description
It was reported that this pacemaker had entered safety mode.Device replacement and product return were recommended.This pacemaker remains in service at this time.No adverse patient effects were reported.Additional information received reported that this pacemaker was explanted and replaced due to entering safety mode.No additional adverse patient effects were reported.This pacemaker was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO MRI DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16738242
MDR Text Key313296940
Report Number2124215-2023-17562
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526558962
UDI-Public00802526558962
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial,Followup,Followup
Report Date 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/09/2018
Device Model NumberL111
Device Catalogue NumberL111
Device Lot Number235240
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 03/31/2023
Initial Date FDA Received04/13/2023
Supplement Dates Manufacturer Received07/26/2023
12/27/2023
Supplement Dates FDA Received08/21/2023
01/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age83 YR
Patient SexFemale
-
-