Model Number V60
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2023
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that the device overheated while in use on a patient.The device was reported to be in use at the time of the reported problem.No patient harm reported.The customer informed the remote service engineer (rse) that the device overheated while in use.The unit was removed from the patient with no reported issues.Th customer confirmed the error code check vent: blower temperature high was in the event log.The rse provided troubleshooting to the customer, advising them to verify the air inlet filter was not dirty or blocked, verify the cooling fan was operational, and then to replace the blower assembly if required.The rse provided the customer with the part information for the blower assembly.In a good faith effort (gfe) response from the customer received, it was confirmed that the advised blower assembly was ordered and replaced.The customer went through the preventative maintenance procedure to verify operation of the device.The customer stated that the replaced part would not be returned, the diagnostic report was deleted during performance verification so it is not available, and that the patient information was not available.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key16741121
MDR Text Key313294991
Report Number2518422-2023-08887
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/17/2023
Initial Date FDA Received04/14/2023
Date Device Manufactured09/08/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-