Catalog Number CV-15122-F
Device Problems Unraveled Material (1664); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2022
Event Type  malfunction  
Event Description
It was reported when the doctor was installing the catheter and when removing it, he saw that the spring wire had frayed.No patient harm was reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
Qn#(b)(4).
 
Event Description
It was reported when the doctor was installing the catheter and when removing it, he saw that the spring wire had frayed.No patient harm was reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
Qn # (b)(4).The actual device was not returned; however, the customer provided one photo for analysis.The complaint of swg unraveled was able to be confirmed by the photo.A complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "warning: do not apply excessive force in removing guide wire or catheters.If withdrawal cannot be easily accomplished, a chest x-ray should be obtained and further consultation requested." without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 20 CM
Type of Device
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16742211
MDR Text Key313741348
Report Number3006425876-2023-00370
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
K862056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCV-15122-F
Device Lot Number71F19M0869
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/21/2023
Initial Date FDA Received04/14/2023
Supplement Dates Manufacturer Received05/10/2023
Supplement Dates FDA Received05/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED; NOT REPORTED
-
-