MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number AZURION 7 B20
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hair Loss (1877)
Event Date 04/04/2023
Event Type  Injury  
Event Description
It has been reported to philips that a patient experienced hair loss on the head after having a procedure on the azurion system.The issue also occurred with another patient which has been reported via philips reference (b)(4) a philips engineer inspected the system onsite and performed system tests.The system dose levels were according to specification and no malfunction of the system was identified.We are conservatively reporting this event as the investigation is ongoing.A follow up report will be submitted when further information is received.
 
Manufacturer Narrative
It was identified that this is a duplicate complaint from an already reported complaint.The investigation will be addressed in mfr report number 3003768277-2023-02314.This complaint will be closed as duplicate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16743097
MDR Text Key313324287
Report Number3003768277-2023-02309
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838085367
UDI-Public00884838085367
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K163715
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 7 B20
Device Catalogue Number722068
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/04/2023
Initial Date FDA Received04/14/2023
Supplement Dates Manufacturer Received07/11/2023
Supplement Dates FDA Received07/19/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-