DENTSPLY IMPLANTS N.V. SURGIGUIDE - 1 IMPLANT; ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS
|
Back to Search Results |
|
Catalog Number UNIQUE REFERENCE PER ORDER (CU |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
Failure of Implant (1924)
|
Event Type
Injury
|
Event Description
|
In this event it is report that while during use of surgiguide on patient the sleeve was to tight and it wouldn't allow the dentist to get to depth on any drills.The fit of the tube was to narrow or tight.It stopped each drill short from where the stopper was.It was all drills starting at the initial drill.The patient was numbed , they did the osteotomy and drilling and that is when they noticed the drills not going to depth, then they tried to place implant and at the point they aborted the surgery due to the issue.They then bone grafted and stitched patient.They are waiting for the patient to heal and possible refer to oral surgery.
|
|
Manufacturer Narrative
|
Therefore, because a serious injury resulted, this event is reportable per 21 cfr part 803.The product planning documentation was checked and did not find any issues.Product return is requested and product will be evaluated after receipt.In case any new or additional information will be gained from this investigation a follow-up report will be sent.Trend is tracked and monitored.
|
|