SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C45091675D-NL |
Device Problems
Leak/Splash (1354); Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Injury
|
Event Description
|
It was reported that the circuit was connected after intubation of the patient and a leak was heard coming from the circuit and were unable to adequately deliver the volumes necessary for the patient.The patient was ventilated with an ambu bag while the circuit was replaced with a new one.No injury was reported.Possible lot numbers reported: 4341247 or 4364557.
|
|
Manufacturer Narrative
|
B3: date of event; d4: udi number are unknown, no product information has been provided to date.D4: expiration date and h4: manufacture date are not available based reported lot number.G5: 510k is blank device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Device available for evaluation codes: updated device evaluation: twenty unused devices, in their original packaging completely sealed, were returned for investigation.Visual inspection and functional testing found no faults or defects.The complaint could not be confirmed.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No corrective action is required as there is no information if the product leaks in the pre-check, which is listed in the instructions for use.
|
|
Search Alerts/Recalls
|
|
|