The lot was manufactured from september 30, 2022 to october 4, 2022.H10: the actual device was received for evaluation.Visual inspection was performed and noted a segment on the tube set had been damaged.A leak was observed at the damaged area of the tube when the sample was leak tested.Indentation marks from the manufacturing flow wrap sealer equipment were observed.The reported condition was verified.The cause of the condition due to an issue with the flow wrap sealer during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|