Catalog Number 2C4009K
Device Problems Fluid/Blood Leak (1250); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter address: (b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume folfusor leaked from the tubing.This was identified when filling the device with 5% glucose.The tubing was melted/crushed a few centimeters from the connection to the device.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The lot was manufactured from september 30, 2022 to october 4, 2022.H10: the actual device was received for evaluation.Visual inspection was performed and noted a segment on the tube set had been damaged.A leak was observed at the damaged area of the tube when the sample was leak tested.Indentation marks from the manufacturing flow wrap sealer equipment were observed.The reported condition was verified.The cause of the condition due to an issue with the flow wrap sealer during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16756155
MDR Text Key313498017
Report Number1416980-2023-01800
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080130
UDI-Public(01)00085412080130
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 05/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4009K
Device Lot Number22J044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/23/2023
Initial Date FDA Received04/17/2023
Supplement Dates Manufacturer Received05/18/2023
Supplement Dates FDA Received05/19/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-