The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
The reported event is unconfirmed as the reported failure could not be reproduced.Photo samples were submitted, however, could not be evaluated for the reported failure.Two extension cables were returned without the original packaging.The cables were labeled 1 - 2 for identification purposes.No damage to the cord was noted.Connections of the cords to the plug and socket were secure.Using a multimeter, the cords were tested for and found to have continuity.The cords were connected to an in-house temperature sensing catheter submerged in a water bath (119118 lot ngevz327) set at 37 degrees celsius, both cords connected to the catheter snuggly with a click and required force to be removed.The cords were then attached to a kilo device, both connected snuggly and required force to be removed.The temperature displayed 37.2 degrees celsius, for both cables, which is similar to the thermometer readings.The samples meet the specification "plug and jack must be firmly secured to wire.As the reported event is unconfirmed, a dhr review is not required.As the reported event is unconfirmed, a label/packaging review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
|