Model Number 214140
Device Problems Difficult to Open or Close (2921); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2023
Event Type  malfunction  
Event Description
It was reported by the healthcare professional in australia that during a shoulder repair procedure on (b)(6) 2023, it was observed that the jaw on the expressew iii suture passer w/o hook device was not closing properly.During in-house engineering evaluation, it was determined that the upper jaw was loose.It was unknown if and how the procedure was completed.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).D4: the expiration date is currently unavailable.Investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of the device received.The device was received and evaluated.Upon visual inspection revealed wear marks on the device as usual in this type of reusables devices.The device does not show any structural anomalies.Upon reviewing the upper jaw, it was found that is loose.When the trigger was fully depressed, the jaw cannot be closed completely.Due to the condition of the jaw, the functional test cannot be performed.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.The manufactured date of this device is 2017 and hence indicates this device is 6 years old.The possible root cause for the loose jaw can be attributed to procedural variables, such handling of the device or product interaction during procedure, a bigger portion of tissue may have been grabbed and forced the jaw to close, also, since this device is reusable, the continuous use and sterilization process can cause metal fatigue leading to a non-closing jaw, however, this cannot be conclusively determined.As per ifu, it is important to inspect the device prior to use to ensure proper mechanical function and do not use if product is damaged.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW III SUTURE PASSER W/O HOOK
Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16762427
MDR Text Key313510530
Report Number1221934-2023-01679
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10886705020218
UDI-Public10886705020218
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number214140
Device Catalogue Number214140
Device Lot Number40515-170119-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/31/2023
Initial Date FDA Received04/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2017
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-