This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and because there was no device malfunction confirmed during evaluation, the definitive root cause of the abnormal flow rate could not be determined.Olympus will continue to monitor field performance for this device.It is possible that the abnormal flow rate may have been caused by detached tubing within the unit, disconnected main board and harness for each valve, faulty electro-pneumatic proportional valve, faulty manifold valve, faulty main board, or faulty first pressure reducer.Olympus will continue to monitor field performance for this device.
|