Catalog Number 121887462 |
Device Problem
Naturally Worn (2988)
|
Patient Problems
Adhesion(s) (1695); Foreign Body Reaction (1868); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Distress (2329); Metal Related Pathology (4530); Unspecified Tissue Injury (4559)
|
Event Date 02/22/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # = > (b)(4).The device catalog number is unknown; therefore, udi is unavailable.Investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Pinnacle metal on metal medical records received.After review of the medial records there was no information provided pertaining to patient harm and device related problem.No revision records received yet doi: (b)(6) 2007, dor: (b)(6) 2023, right hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
After a review of medical records patient was revised due to metal-on-metal hip failure.Operative notes indicated there was sort of typical brown almost sawdust-type material present in the hip joint.There were some corrosion products on the trunnion itself and these were cleaned off.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.Added: d4 (lot, catalog, udi), g4 (pma/ 510(k)), h4 corrected: d1, d2a, d10 concomitant.
|
|