SMITH & NEPHEW, INC. MI Z HANDLE ACET REAMER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71364073 |
Device Problems
Break (1069); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/27/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that, during a thr, the internal mechanism of one (1) mi z handle acet reamer broke during reaming.No pieces fell into the patient.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not harmed as consequence of this problem.
|
|
Manufacturer Narrative
|
H10: internal complaint reference: case (b)(4).
|
|
Manufacturer Narrative
|
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device reveals the inner shaft broke into two pieces.All pieces of the device were returned.The device shows signs of significant wear and use.A review of complaint history based on the historical data revealed a similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that a similar event was previously identified, and prior actions were performed.In addition, it has been concluded that there is a potential for breakage if used after the expected lifetime or excessive force is used as this device is a reusable instrument and it is expected to wear over time.Also, the failure rate of this issue is within expected and documented in the risk file.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|