Model Number 71364073
Device Problems Break (1069); Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/27/2023
Event Type  malfunction  
Event Description
It was reported that, during a thr, the internal mechanism of one (1) mi z handle acet reamer broke during reaming.No pieces fell into the patient.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not harmed as consequence of this problem.
 
Manufacturer Narrative
H10: internal complaint reference: case (b)(4).
 
Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device reveals the inner shaft broke into two pieces.All pieces of the device were returned.The device shows signs of significant wear and use.A review of complaint history based on the historical data revealed a similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that a similar event was previously identified, and prior actions were performed.In addition, it has been concluded that there is a potential for breakage if used after the expected lifetime or excessive force is used as this device is a reusable instrument and it is expected to wear over time.Also, the failure rate of this issue is within expected and documented in the risk file.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MI Z HANDLE ACET REAMER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16781930
MDR Text Key313743058
Report Number1020279-2023-00888
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010507570
UDI-Public03596010507570
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364073
Device Catalogue Number71364073
Device Lot Number21KM08735
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2023
Initial Date FDA Received04/20/2023
Supplement Dates Manufacturer Received05/11/2023
Supplement Dates FDA Received05/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-