As reported, during use in patient with this pressure monitoring set, incorrect diastolic (20-25mmhg) and slightly underestimated systolic (100-110mmhg) measurements were obtained.The waveform obtained was normal.Initially, this was corrected recalibrating the device.Shortly afterwards, the blood pressure value was underestimated again, which led to an increase in the noradrenaline infusion (0.5 ug/kg/min).Finally, a new "zero" calibration revealed severe iatrogenic hypertension due to the excessive increase in noradrenaline.After this, the device was replaced and the issue was solved.This episode lasted 20 minutes.No further treatment was required to correct the hypertension, and there was no allegation of patient injury.The device was not available for evaluation.
|