An olympus field service engineer (fse) was dispatched to evaluate and repair the device.The fse noted good water pressure and that the filters were recently changed, air was noted in the water line and the fse purged the air, fse cleaned and inspected the machine, checked disinfectant, detergent and alcohol, software attributes were verified and confirmed, and the equipment was repaired and verified according to original equipment manufacturer (oem) instructions.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
This supplemental report is to inform that upon further review, this is not a reportable malfunction.The initial medwatch reported the subject device exhibited error e01 and e73 indicating supply water irregularities; however; per the legal manufacturer, this issue is not likely to cause or contribute to death or serious injury if the malfunction were to recur.
|