During the end of a procedure, an operating room staff member smelled a smoke-like odor that was identified as overheating of the outlets on a relocatable power tap (rpt) located next to the operating room table.Patient was unaffected; no flame or fire.The rpt(radiopharmaceutical therapy) was mounted on an equipment assembly next to the operating room table.Upon smelling, then seeing smoke coming from the outlets, the assembly was moved away from the table, all devices were unplugged, and the rpt(radiopharmaceutical therapy) was unplugged from the all outlet.Rpt(radiopharmaceutical therapy) was removed from service and later recovered by facilities staff member.Investigation found that the outlet overheated due to user error: power of the rpt(radiopharmaceutical therapy) output was exceeded; high power requirement item was plugged incorrectly.
|