The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.The main power switch was found jammed (depressed) and non-operational during the device evaluation.Additionally, the lamp life meter was expired, with over 500 hours of use.The output socket was found worn-out and was compromising the scope connection.The front panel was damaged.The chassis was worn-out.The base plate was corroded, and the device was heavily soiled on the inside.If additional information becomes available following the device evaluation, a supplemental report will be filed.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Correction to d4 (udi number) and h4, information was inadvertently not included on the initial medwatch.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, it is likely the power switch not able to be turned on occurred due to a power switch failure.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
|