• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BUNNELL, INC. LIFEPULSE; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BUNNELL, INC. LIFEPULSE; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 204
Device Problems Circuit Failure (1089); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2023
Event Type  malfunction  
Event Description
While patient was out of bed with mother, jet alarmed "high level" referring to the water level in the heater.Patient was removed from jet ventilator and provided ventilation with t-piece resuscitator.Jet ventilator was put through a system test without resolution of alarm.Jet was power cycled and put through a second system test without resolution.Original circuit removed and new circuit placed, alarm resolved.Prior to placing patient back on jet ventilator, entire circuit flooded with water.Jet was removed from service and tagged, patient was placed on new jet ventilator without complication.Patient remained stable for duration.No patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPULSE
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
BUNNELL, INC.
436 lawndale dr.
s salt lake UT 84115
MDR Report Key16806196
MDR Text Key313935655
Report Number16806196
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number204
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/12/2023
Event Location Hospital
Date Report to Manufacturer04/25/2023
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/25/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age5 DA
Patient SexFemale
-
-