• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD20 BIPLANE
Device Problem Radiation Underexposure (3018)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2023
Event Type  malfunction  
Event Description
It has been reported to philips that the system displayed an error message - test shot underexposed.The issue was found during clinical use.The operation stopped and resulted in a delay to therapy.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information acquired, the system was in clinical use for a planned neurovascular procedure when the system displayed a "test shot underexposed" messaged during dsa (digitally subtracted angiography).The procedure was completed with a delay of 15 minutes.A philip remote service engineer (rse) analyzed the issue remotely and identified that the message was due to fluoroscopy being active for less than 3 seconds before dsa was started.No malfunctions of the system were identified via log file analysis.The system was returned to use in good working order.At the time this complaint was received, philips did not have enough information to exclude the potential for death or serious injury on recurrence, and as such the complaint was reported.Since that time, new information has been received which has led to the determination that this is scenario is not likely to cause or contribute to death or serious injury if it were to recur.The message displayed by the system was related to a workflow issue and no malfunction has been identified.Based on the investigation results, philips concluded that the complaint is not reportable.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16807466
MDR Text Key313926619
Report Number3003768277-2023-02500
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059061
UDI-Public00884838059061
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20 BIPLANE
Device Catalogue Number722013
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/18/2023
Initial Date FDA Received04/25/2023
Supplement Dates Manufacturer Received09/07/2023
Supplement Dates FDA Received12/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-