MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS WITH FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS WITH FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7361-24
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2023
Event Type  malfunction  
Event Description
It was reported that disposable came loose.No patient injury was reported.
 
Manufacturer Narrative
No product was returned.We are unable to confirm the reported complaint.If the product is returned, smiths medical will reopen this complaint for further investigation.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product was returned; therefore, no visual and functional tests were performed, the reported complaint could not be confirmed, and the root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SETS WITH FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16809259
MDR Text Key313948493
Report Number3012307300-2023-04512
Device Sequence Number1
Product Code FPA
UDI-Device Identifier15019517191738
UDI-Public15019517191738
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
K031361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number21-7361-24
Device Catalogue Number21-7361-24
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/05/2023
Initial Date FDA Received04/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-